Saltar a la búsqueda Saltar al contenido principal Saltar a la navegación

United Kingdom 0800 488 488 | Intl +44 (0) 161 777 6611 | Welcome to our new website, found an issue or bug? Please report it here

Hemos notado que estás en , por favor visita tu sitio local haciendo clic en el enlace a continuación para ofertas y soporte regionales.

CL309A-Calibrator

Calibrador portátil de bucle de corriente de 4 a 20 mA con alimentación de bucle de 24 V

Handheld 4 to 20 mA Current Loop Calibrator with 24 V Loop Power
CL309A-Calibrator
Desde SGD 2,132.87
Handheld 4 to 20 mA Current Loop Calibrator with 24 V Loop Power CL309A-Calibrator CL309A-Calibrator
loading all models
Resumen del Producto
  • Transmisor de 2 hilos Simular Carga de voltaje (Transmisor de 2 cables): 2 V a 20 mA; Protección contra sobrecarga/límite de corriente (Transmisor de 2 cables): 24 mA; Límites de voltaje del lazo (Transmisor de 2 cables): 2 a 100 Vdc (protegido sin fusible contra conexiones de polaridad inversa); Vida útil de la batería (Transmisor de 2 cables): 125 h
  • Duración de la batería 30 h
  • Tamaño y tipo de batería Alcalina AA
  • Altura 4.06 cm
  • Longitud 14.3 cm
  • Límites de suministro del bucle Tensión de cumplimiento: 24 Vcc @ 20 mA; Capacidad de conducción: 1200 Ω a 20 mA durante 15 horas nominales
  • Nota sobre la exactitud de la medición Precisión (Fuente): Comprobaciones EZ a 4 y 20 mA(*) ±0,025 % a 4,00 mA y 20,00 mA (±0,005 mA) 0,0 a 24,00 mA ±0,05 % de 24,00 mA (±0,01 mA); Precisión (Lectura mA): Precisión: Por debajo de 24,01 mV: ±0,05 % de 24,00 mA (±0,01 mA) Por encima de 24,00 mV: ±0,05 % de 52,00 mA (±0,02 mA); Precisión (Lectura de voltaje): ±0,05 % de FS; Efecto de la temperatura (Lectura de voltaje): ±100 ppm/°C de FS
  • Rango de medición Rangos y Resolución (Fuente): 0,00 a 24,00 mA rango completo o -25,0 a 125,0 % de 4 a 20 mA; Corriente (Lectura mA): 4 a 20 mA; Nota de Corriente: 0,00 a 52,00 mA rango completo O -25,0 a 300,0 % de 4 a 20 mA; Caída de Voltaje (Lectura mA): 2V a 50 mA; Vida de Batería (Lectura mA): 125 h; Rangos (Lectura de Voltaje): -99,99 a 99,99 Vdc rango completo (FS); Resistencia de Entrada (Lectura de Voltaje): >2 MΩ; Vida de Batería (Lectura de Voltaje): >125 h
  • Varios Indicación de batería baja con aproximadamente 1 hora de funcionamiento restante; protección contra sobretensión hasta 135 vrms (clasificada para 30 segundos) o 240 vrms (clasificada para 15 segundos); pantalla gráfica de cristal líquido de alto contraste con dígitos de 10,5 mm (0,413") de altura
  • Ruido ± 0,005 mA de lectura
  • Temperatura de operación, máx. 60 °C
  • Temperatura de operación mínima -20 °C
  • Protección contra sobrecorriente 24 mA CC
  • Humedad relativa 10 % ≤HR ≤90 % (0 a 35 °C), sin condensación, 10 % ≤HR ≤70 % (35 a 60 °C), sin condensación
  • ID de serie CL309A-Calibrator
  • Temperatura de almacenamiento, máx. 60 °C
  • Temperatura de almacenamiento, mínima -30 °C
  • Alimentación Nota 4 pilas alcalinas “AA” (incluidas) 1,5 V (Duracell MN1500 o equivalente); adaptador de corriente alterna opcional de 120 Vca 50/60 Hz disponible; kit de batería recargable NiMh disponible
  • Efecto de la temperatura ± 0,005 %/°C de FS (*) Estos son puntos de referencia cardinales calibrados internamente y fuente/potencia y transmisor de medida de 2 cables
  • Peso 12.1 oz
  • Ancho 7.62 cm

Supported Applications

The Omega CL309A is a handheld current loop calibrator designed to check, calibrate, and measure instruments in a 4 to 20 mA DC loop. The device sources and reads current signals, simulates a 2-wire transmitter, and powers 2-wire transmitters while measuring their output. It displays current readings in milliamps or as a percentage of the 4-20 mA range.

Operating Conditions & Performance

The CL309A operates within an ambient temperature range of -20 to 60 °C (-5 to 140 °F) and supports storage from -30 to 60 °C (-22 to 140 °F). Relative humidity limits are defined as 10% ≤RH ≤90% (0 to 35 °C), Non-condensing, or 10% ≤RH ≤70% (35 to 60 °C), Non-condensing. Accuracy specifications vary by mode: the device achieves ±0.025% at 4.00mA and 20.00mA (±0.005mA) for source, power measure, and simulate functions. For EZ-Check™ settings at 4 & 20 mA, accuracy is ≤ ± 0.05% of 24.00mA (±0.01mA). In read modes below 24.01mA, accuracy is ≤ ± 0.05% of 24.00mA (±0.01mA); above 24.00 mA, it is ≤ ± 0.05% of 52.00mA (±0.02mA). Voltage read accuracy is ≤ ± 0.05 % of Full Scale with a resolution of 10mV.

Performance includes loop compliance voltage ≥ 24 DCV @ 20.00mA and loop drive capability of 1200 Ω at 20 mA for 15 hours nominal. The device features overload/current limit protection up to 54 mA nominal in simulate modes, with a voltage burden ≤ 1.2V at 20 mA (simulate) or ≤ 2V at 50 mA (read). Battery life is ≥ 30 hrs at 12 mA nominal for source and power measure mode, and ≥ 125 hours nominal for simulate and read modes.

Configuration Options

The CL309A series supports the following measurement ranges and configurations:

  • Current Ranges: Source/Power Measure/Simulate 0.00 to 24.00 mA Full Span OR -25.0 to 125.0% of 4-20 mA; Read mode supports 0.00 to 52.00 mA Full Span OR -25.0 to 300.0% of 4-20 mA.
  • Voltage Range: Reads DC volts from -99.99 to +99.99 VDC with input resistance ≥ 2 MΩ.
  • EZ-CHECK™ Settings: Pre-set outputs at 4.00 mA (0.0%) and 20.00 mA (100.0%).
  • Power Supply: Operates on four (4) 'AA' Alkaline 1.5V batteries (Duracell MN1500 or equivalent).
  • Connections: Red and black banana plugs.
  • Housing: Includes a rubber boot with overall dimensions L=5.63 x W=3.00 x H=1.60 inches (L=14.30 x W=7.62 x H=4.06 centimeters) and weighs 12.1 ounces including boot & batteries.

Optional accessories include a (100 to 120) VAC Adaptor, a (200 to 240) VAC Adaptor, and a NiMH Rechargeable battery kit. The display is a high contrast graphic liquid crystal with 0.413" (10.5 mm) high digits.

Key Product Differences

The CL309A model functions as an integrated handheld unit combining source, measure, and simulate capabilities in a single enclosure. The device distinguishes between Source/Power Measure modes (limited to 24 mA full span) and Read mode (extended to 52 mA). It provides specific EZ-Check™ presets for rapid verification at standard loop points.