OS310-Series

Thermomètre infrarouge haute performance de -20 à 1650 °C, 50:1

Modèles en stock
  • -50 to 1650°C (-58 to 3002°F) temperature range
  • 50:1 Field of View
  • Adjustable Emissivity: 0.10 to 1.00
  • 250 ms response time
OS310-Series
Depuis C$ 463.18
Modèles en stock
Thermomètre infrarouge haute performance de -20 à 1650 °C, 50:1 OS310-Series OS310-Series
loading all models
L'aperçu du produit
  • Précision ± 1.5 °C (± 2.7 °F)
  • Détail de la précision -50 to 0°C (-58 to 32°F): ± 3°C (± 5°F) 0 to 1650°C (32 to 3002°F): ± 1.5% or 1.5°C (± 2.7°F)
  • Alarm High and Low
  • Battery Size and Type Alkaline 9V
  • Display Monitoring Function Average, Max, Memory, Min
  • Display Type LCD
  • Emissivity, Max 1
  • Emissivity, Min 0.1
  • Field of view 50:1
  • Form factor Pistol Grip
  • Height 200 mm
  • Laser Dot
  • Length 60 mm
  • Operating Temperature, Max 40 °C
  • Operating Temperature, Min 0 °C
  • Process Temperature, Min -50 °C
  • Relative Humidity, Max 0.95
  • Relative Humidity, Min 10%
  • Relative Humidity Note 10 -95% RH non-condensed
  • Repeatability <1/3
  • Repeatability Note Less than 1/3 of basic error
  • Response Time <250 ms
  • Series ID OS310-Series
  • Spectral Response, Max 14 µm
  • Storage Temperature, Max 50 °C
  • Storage Temperature, Min -20 °C
  • Weight 270 g
  • Width 141 mm

The OS310-series handheld digital IR thermometer is used for non-contact temperature measurement of metals, ceramics, graphite, etc. in temperature ranges between -50°C (-58 °F) and 1650°C (3002°F). This portable battery powered unit has a 50:1 field of view, displays current temperature as well as minimum, maximum, average and differential temperatures. The large 50:1 field of view is perfect for hazardous applications where the user can't get close to the target area or for applications where the target size is smaller but far away. The OS310-Series additional features include a backlit LCD display, adjustable emissivity, alarm settings and a single laser dot to help locate the target.


Accuracy Chart

Device Accuracy = ± 1.5% rdg or 1.5°C, whichever is greater

Target Temperature

Upper Limit Reading

Lower Limit Reading

5°C

6.5°C

3.5°C

100°C

101.5°C

98.5°C

150°C

152.25°C

147.75°C

400°C

406°C

394°C

1200°C

1218°C

1182°C

  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 3ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;

Any way to get the default temp to read in Fahrenheit not Celsius on this model? It keeps restarting in Celsius..

Asked by: Drewzilla
Hi Drewzilla, Thank you for the inquiry. We will respond again shortly. Verifying power up state, especially Fahrenheit selection. Best regards
Answered by: Norwalk AE_RDS
Date published: 2025-12-10
  • y_2025, m_12, d_30, h_8CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.42
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_CA, sid_OS-310-501, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 4ms
  • QUESTIONS, PRODUCT
fr-CA
fr
Enregistrer la configuration