Product Overview
- Accuracy ± 1 %
- Additional Features 1 1/4" DIN, Sensors Sold Separately
- Altitude,Max 3000 m
- Conductivity Measurement 0 to 999.9 μS/cm, 0 to 9999 μS/cm, 0 to 100mS/cm, 0 to 600 mS/cm
- Cutout Dimensions 92 × 92 mm
- Display Size 13 mm (0.51 in)
- Display Type LCD
- Enclosure Ratings IP54
- Frequency 50/60 Hz
- Height 96 mm
- Impedance 750 Ω
- Length 110 mm
- Mounting Orientation Panel Mount
- Number of Relays 2
- Operating Humidity 5 to 95% RH
- Operating Range Note Resistivity: 0 to 18 MΩ or 0 to 19.99 μS/cm
- Operating Temperature, Max 55 °C
- Operating Temperature, Min -10 °C
- Output Detail Type: Isolated 4 to 20 mA Range: Selectable zero and span Accuracy: ±0.02 mA
- Process Temperature, Max 0 °C
- Process Temperature, Min 100 °C
- Relay Note 2 programmable , 10 A Control Type: On/off Relay Set Point Hysteresis: User programmable
- Resolution 0.01 to 0.01 mS/cm; Temperature: 0.1°C (32.1°F)
- Temperature Accuracy ± 0.2 °C
- Weight 650 g
- Width 96 mm
Industrial process conductivity/temperature controllers CDTX-111 and CDTX-112 are panel instruments for in-process monitoring of conductivity. Range switch over and cell constant checks can both be freely set and adjusted during operation. Unique signal collection and processing technology, and a special thick film circuit are employed to ensure accurate measurements and stable operation. Other features include linearized data, automatic temperature compensation, and maintenance-free cells. The control functions trigger relays based on conductivity set-points. The CDTX-111 and CDTX-112 are ideal auxiliary instruments for many types of water testing applications.