1-800-663-4209 | Bienvenue sur notre nouveau site Web, vous avez rencontré un problème ou un bug? Veuillez le signaler ici
ISD-TC-Series
Surveillance de la température basée sur le web
- Enregistreur de graphiques virtuel
- Serveur Web
- Double thermocouple (10 types)
- Alarmes par courriel ou message texte
ISD-TC-Series
À partir de
C$
1,362.24
Aperçu du produit
- Température de fonctionnement de l'adaptateur secteur 0 à 40 °C
- Alarmes Deux entrées de contact, TTL 0,5 mA avec pull-up 10K ; une sortie collecteur ouvert 150 mA @ 30 Vcc
- Homologations CE, FCC-B
- Température de fonctionnement de la batterie -18 à 55 °C
- Taille et type de batterie Alcaline 9 V
- Température maximale du câble 125 °C
- Température minimale du câble -40 °C
- Configuration Interface Web interne
- Interface de données Ethernet
- Remarque sur l'interface de données Ethernet : (RJ45) Fixe ou négocié automatiquement 10/100BASE-T, AUTO MDI/MDIX Protocoles : TCP, UDP, SNMP, SMTP, NTP, ARP, ICMP, DHCP, DNS, HTTP et Telnet
- Affichage LCD
- Remarque sur l'affichage 32 chiffres 4,8 x 9,7 mm (0,19 x 0,38")
- Facteur de forme Montage sur support mural
- Hauteur 1.31 in
- Type d'entrée Externe
- Longueur 3.47 in
- Type de mesure Thermocouple
- Mémoire Carte Flash SD 2 Go
- Remarque sur la mémoire Carte 2GB : 8 mois de stockage de données avec des intervalles d’enregistrement de 1 seconde ou 7 ans avec des intervalles de 10 secondes
- Nombre de mesures 2
- Nombre de relais 2
- Température de fonctionnement, max 60 °C
- Température de fonctionnement, min 0 °C
- Sorties relais 1,5 A à 30 Vcc
- Échantillons, max 85 000 échantillons, intervalle de 16 échantillons/seconde, intervalles d’enregistrement de 1 seconde ou 7 ans à des intervalles de 10 secondes
- Type du capteur Thermocouple
- ID de série ISD-TC-Series
- Logiciel Serveur OPC ; macro pour l'enregistrement des données dans le programme Excel ; compatible avec les systèmes d'exploitation Windows®
- Température de stockage, max 85 °C
- Température de stockage, min -40 °C
- Approvisionnement Remarque Adaptateur d’alimentation CA qualifié pour la sécurité :, Entrée : 100 à 240 VCA, 50/60 Hz, Sortie nominale : 9 VCC à 0,5 A, (inclus), Batterie de secours : 9 VCC, alcaline (incluse)
- Type d'approvisionnement électrique CC/Batterie
- Tension d'alimentation 9 à 12 Vcc
- Stabilité de la température 0,08°C/°C
- Suivi de l'extrémité froide du thermocouple 0,05°C/°C
- Résistance du fil du thermocouple 100 Ω max
- Poids 1.02 lb
- Largeur 5.13 in
Surveillance de la température sur le Web La série ISD-TC offre un moyen pratique et efficace de suivre à distance les données de température via une plateforme en ligne. Ce système permet aux utilisateurs d'accéder en temps réel aux relevés de température depuis n'importe quel endroit disposant d'une connexion Internet, garantissant ainsi une surveillance continue et des alertes en temps opportun. Idéal pour les applications nécessitant un contrôle précis de la température, il améliore la sécurité, la conformité et l'efficacité du fonctionnement en offrant des fonctionnalités faciles de visualisation, d'enregistrement et de rapport des données.
PDF & Manuels
Afficher les évaluations et les avis
Rated 3 out of
5
by
Rider from
Solid product, but complicated to set up
We recently purchased the iSD-TC for remote temperature monitoring and email alerts for laboratory processes. The product is well made and durable. Temperature readouts are consistent and clear. However, we're still trying to get it to connect directly to a computer or directly to the internet. Contacted customer support, and they just gave me a link to the manual. It's definitely not plug-and-play. If we could get it to connect, upload, and give results within a day after receiving, this review would be 5 stars.
Date published: 2017-10-16
Rated 1 out of
5
by
EngineerIsaac from
Extremely unreliable
We have 3 of these units on our drug coolers and have now had to throw away several lots worth over $100k because their OS freezes up randomly when attached to a network. Unfortunately it doesn't give you a warning when it freezes up, it just stops recording critical data. We ordered brand new units thinking it would fix the problem and the new unit's OS froze up within two days. Hope they fix this bug as it would be a decent product otherwise. Their tech support was not helpful whatsoever.
Date published: 2022-02-04
Rated 2 out of
5
by
Isaac99 from
Would be a great product but very buggy.
I've had nothing but headaches with the three units we have onsite. They regularly freeze up and lose communication. When they freeze up they still display "Data Recording" but are not actually recording data which resulted in our company having to throw away thousand of dollars in product. I hope they work out the bugs and they might be good units.
Date published: 2021-04-15
Rated 4 out of
5
by
thinman from
Great, but replacing the battery is a hassle
Been buying these for years for our minus 80 freezers since it is one of the few units that can read temperatures so low. My only complaint is that the unit has to be taken apart in order to replace the 9v backup battery, which has to be done about every 8 months. It would really improve the product if that was fixed. Tech support has always been helpful too.
Date published: 2020-05-09
Rated 5 out of
5
by
mtc8 from
We have over 10 of these, Great Product
I've installed these over a year ago and they have paid for themselves over and over. There are 2 ways to connect to these to get data, but i prefer the Omega Enterprise Gateway which is able to display all on one page. The other way is to connect to the IP address of the monitor. Both work, alarms are invaluable. Great Product, came back here because i'm buying more.
Date published: 2019-04-15
Rated 1 out of
5
by
GEHC from
Unable to obtain data
Just like the previous comment, I am unable to download the data as well. Please advise a fix.
Date published: 2018-07-30
how can i calibrate and write offset values or table to this instrument. Omega i server , ISD-TC
Please direct your question to our DAS Engr dept -DAS@OMEGA.COM
Date published: 2020-12-09
where can i find a upgrade for firmware?
Thank you for your question. We will require more information to help on this question, please email the DAS team at DAS@OMEGA.COM
Date published: 2022-08-04
Is a battery required if I don't care about recording if power is disconnected? Will the unit still function normally w/o a battery installed?
Yes, ac power only is ok
Date published: 2020-06-18
Does the isd-tc have a count down timer that freezes the unit when trying to make entries online?
Thank you for your question. There is no count down timer that will freeze the unit while making entries.
Date published: 2022-08-18
What is the part number of the Power Adapter? I need to do NRTL certification and have to provide this information for that testing.
Hello John
Replacement power supply part number is UNIV-AC-100/240-12-5A
Date published: 2021-03-16