Product Specs
<< See all Product ModelsBrand |
![]() |
Accuracy | ± 1 % |
Accuracy Detail | ± 1% of FS at calibration temperature and pressure |
Altitude Sensitivity | Deviation of up to 1% from stated accuracy, after re-zeroing |
Calibration | NIST 10-Point Certificate |
Compliance | CE |
Data Interface | RS232 |
Display Type | None |
Electrical Output | 4 to 20 mA, 0 to 5 Vdc |
Enclosure Ratings | Inst. Level II/Poll. Degree II (IEC 60664) |
Humidity Range | Up to 70% |
Leak Rate Under Vacuum | less than 10^⁻9 cc He/s |
Materials, Wetted | 316 Stainless Steel, Aluminum, Brass, FKM |
Media Compatibility | Air, Gases, Nitrogen, Oxygen |
Operating Pressure, Max | 500 psi |
Operating Pressure, Min | 0 psi |
Operating Temperature, Max | 50 °C |
Operating Temperature, Min | 5 °C |
Output Detail | Linear 0 to 5 Vdc (3000 Ω min load impedance); Linear 4 to 20 mA (500 Ω max loop resistance). Max noise 20 mV peak to peak (for 0 to 5 Vdc output) |
Output Signal | Current, Voltage |
Power Consumption | 2.4 W |
Pressure Coefficient | 0.01% FS/psi |
Pressure Drop, Max | 0.18 psi |
Process Connection Size | 1/4 in |
Process Connection Type | Compression Fitting |
Repeatability | ± 0.15% |
Response Time | 2 sec |
Response Time Note | 600 ms time constant; approximately 2 seconds to within ± 2% of set flow rate for 25 to 100% of full scale flow |
Supply Power Note | Transducer Input Power: 11 to 26 Vdc, 100 mV max peak to peak output noise |
Supply Voltage | 11 to 26 Vdc |
Switch Rating | 30 Vdc, 1A |
Switch Type | SPDT NO/NC, Selectable |
Temperature Coefficient | 0.15% FS/°C |
Turndown Ratio | 50:1 |
Volumetric Flow Rate, Max | 20 SCCM |
Volumetric Flow Rate, Min | 0.4 SCCM |
See All Specs |