Skip to search Skip to main content Skip to navigation

1-800-663-4209 | Welcome to our new website, found an issue or bug? Please report it here

We noticed you are from . Visit your local site for regional offers and live support.

FMA6000-Series

Multi Parameter Mass Flow Meter With Flow Totalizer

  • Multi-Drop Capability of Up to 256 Units (for RS-485 Option)
  • Stores Calibration Data for Up to 10 Different Gases
  • Supports 10 Different Engineering Units Including User Defined
  • Programmable 12 Digits Totalizer Indicates Total Gas Volume
FMA6000-Series
From $ 1,542.00
Multi Parameter Mass Flow Meter With Flow Totalizer FMA6000-Series FMA6000-Series
loading all models
Product Overview
  • Accuracy ± 1 %
  • Accuracy Detail 0°C to 50°C and 5 to 100 psia (34.5 - 689.5 kPa): ± 1% of full scale (F.S.)
  • Additional Features 1 2 SPDT 1A Relay, Multi-parameter Display
  • Altitude Sensitivity Incremental deviation of up to 1% from stated accuracy, after re-zeroing
  • Calibration NIST 10-Point Certificate
  • Circuit Protection Reverse polarity, Current Limit, Resettable Fuse
  • Compliance CE
  • Data Interface RS485
  • Data Interface Note The digital interface operates via RS485 (optional RS232 is available)
  • Display Note 128 x 64 graphic LCD with backlight (up to 8 lines of text). User selectable backlight saver.
  • Display Type LCD
  • Electrical Output 4 to 20 mA, 0 to 5 Vdc, 0 to 10 Vdc
  • Enclosure Ratings Inst. Level II/Poll. Degree II (IEC 60664)
  • Humidity Range Up to 70%
  • Leak Rate Under Vacuum less than 10^⁻9 cc He/s
  • Materials, Wetted 316 Stainless Steel, 416 Stainless Steel, FKM
  • Media Compatibility Air, Gases, Nitrogen, Oxygen
  • Operating Pressure, Max 100 psi
  • Operating Pressure, Min 0 psi
  • Operating Temperature, Max 50 °C
  • Operating Temperature, Min -10 °C
  • Output Detail Linear 0-5 Vdc (3000 ohms min load impedance); Linear 0-10Vdc (6000 ohms min impedance); Linear 4-20 mA (500 ohms maximum loop resistance). Maximum noise 20mV peak to peak (for 0-5 Vdc output).
  • Output Signal Current, Voltage
  • Pressure Accuracy ± 1 %
  • Pressure Coefficient 0.01% FS/psi
  • Pressure Drop, Max 0.04 psi
  • Process Connection Type Compression Fitting
  • Process Temperature, Max 50 °C
  • Process Temperature, Min 0 °C
  • Repeatability ± 0.15%
  • Response Time 300 ms
  • Response Time Note 300ms time constant; approximately 1 second to within ± 2% of set flow rate for 25% to 100% of full scale flow.
  • Supply Power Note May be configured for three different options: Bipolar ± 15Vdc (± 200 mA maximum); Unipolar +12Vdc (300 mA maximum); Unipolar +24Vdc (250 mA maximum); Circuit boards have built-in polarity reversal protection. Resettable fuses provide power input protection
  • Supply Voltage Bipolar ± 15Vdc, Unipolar 12Vdc to 24Vdc
  • Temperature Accuracy ± 1 °C
  • Temperature Coefficient 0.15% FS/°C
  • Volumetric Flow Rate, Min 0.2 SCCM
  • Weight 2.2 lb

The Multi Parameter Mass Flow Meter with Flow Totalizer offers precise flow measurement with ±1% accuracy and ±0.15% repeatability. It supports operating pressures up to 100 psi and features a fast 300 ms response time. Calibration is certified with a NIST 10-point certificate. The device includes a local keypad, LCD interface, and digital communication via RS485 or optional RS-232. Compatible with air, gases, nitrogen, and oxygen, it provides multiple electrical outputs and meets CE compliance.

Features

All features of the FMA6600 and FMA6700 Series flow meters can be accessed via the local four button keypad and LCD. The digital interface operates via RS485 (optional RS-232 available) and provides access to applicable internal data including: flow, temperature, pressure reading, auto zero, totalizer and alarms settings, gas table, conversion factors and engineering units selection, dynamic response compensation and linearization table adjustment. The analog interface provides 0 to 5 Vdc or 0 to 10 Vdc or 4 to 20 mA outputs for flow, pressure and temperature (jumper selectable).


Auto Zero
The FMA6600 and FMA6700 Series supports automatic sensor zero offset adjustment that can be activated locally via keypad or remotely via digital interface. The auto zero feature requires absolutely no flow through the meter during auto zero process. Provisions are made to either start, stop or save the current auto zero value via digital commands.


Totalizer
The total volume of the gas is calculated by integrating the actual gas flow rate with respect to time. Both keypad menu and digital interface commands are provided to: • set the totalizer to ZERO • start the totalizer at a preset flow • assign action at a preset total volume • start/stop totalizing the flow • read totalizer Totalizer conditions become true, when the totalizer reading and the "Stop at Total" volumes are equal.

Accessories

cables

25-PIN D CONN/6FT REPL CNTRLR
$111.26
2 In Stock

meters/controllers

115 Vac to 12 Vdc power supply
$88.54
0 In Stock
This item is discontinued. Please contact Technical Sales Team for assistance.
230 Vac to 12 Vdc power supply
$102.47
0 In Stock
This item is discontinued. Please contact Technical Sales Team for assistance.
239 Vac to ±15 Vdc power supply
$309.99
0 In Stock
This item is discontinued. Please contact Technical Sales Team for assistance.
230 Vac to 24 Vdc power supply
$109.94
0 In Stock
This item is discontinued. Please contact Technical Sales Team for assistance.
115 Vac to 24 Vdc power supply
$94.98
0 In Stock
This item is discontinued. Please contact Technical Sales Team for assistance.
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 9ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 16ms
  • QUESTIONS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
en-US
en
Save Configuration