Microprocessor controlled 2-wire 4 to 20 mA transmitter.Converts a low level, frequency signal from a flow sensor into an analog 4 to 20 mA output.
$1,408.07
2 In Stock
Show Ratings & Reviews
Is there somewhere I can download a 3D model of this flow meter?
Thank you for your inquiry and unfortunately, Omega does not provide CAD or STEP files for their products.
Date published: 2021-03-18
I need to connect the FTB-102 with a PLC. Can this flowmeter generate a 4-20mA or a DC signal or even a digital pulse output which can be read on the controller to monitor flow through the meter?
Thank you for your inquiring about this this flow meter. The FTB-100 series turbine flow meter self generates a sinusoidal wave and will need an amplifier such as the FLSC-AMP to amplify to a square pulse.
Date published: 2020-10-01
Are these good for zone 0?
Thank you for your inquiry in regard to the turbine flow meter. The FTB-100 series coupled with the signal conditioner identified in the data sheet is ATEX rated, without it, unfortunately, no.
Date published: 2020-04-27
Do any of the pins connect with the body of the flow meter or housing the signal conditioner sits in?
Thank you for your inquiry about this flow meter. The turbine flow meter and signal conditioner are recommended to be purchased as a 'SYS' or system and they are scaled together to be used when received.
Date published: 2020-06-04
If the system with the signal conditioner is rated for Zone 0, why does the datasheet say the signal conditioner is only rated for Zone 1 and 2? Thanks,
Thank you for your inquiry. The system is only rated for Zone 1 and 2, not Zone 0.
Date published: 2020-05-06
Viscosity affects the reading? I.e. a sensor calibrated for water will have same readings when used for oil with viscosity between 25 to 75 cSt (viscosity changing by oil temperature) ?
Thank you for your inquiry about the turbine flow meter. Yes, the viscosity affects the reading and the turbine flow meter is calibrated for water and if you have a viscous liquid to measure, then, it will need to be specifically calibrated for that liquid for an additional cost as noted on the data sheet. Please note, in the event the temperature changes so does the viscosity and therefore, the calibration will be affected as a result. Highly recommend a positive displacement flow meter such as the FPD3000 series.
Date published: 2021-06-01
Do any of the electrical connections have an electrical connection with the housing? Is it isolated? I'm using with a piece of equipment that requires an SELV power supply and equipment. The housing will be PE grounded.
The self powered flow meter does not come with an electrical housing. The body of the flow meter needs to be grounded if used in a noisy environment.
Date published: 2020-06-03
If the flow meter with the signal conditioner is good for Zone 0 as is claimed be the previous answer, why does the data sheet say the signal conditioner only good for Zone 1 and Zone 2?
According to our specifications the signal conditioner is not rated for Zone 0. Only good for zone 1 and 2.
Date published: 2020-05-06