1-800-663-4209 | Welcome to our new website, found an issue or bug? Please report it here
HH376-Series
1 Channel High Precision Handheld RTD with Probe Datalogger
Models In Stock
- High Accuracy with 0.01C/F Resolution
- Large Backlit Dual LCD Display
- MAX/MIN/HOLD Function
- 16,000 Record Data Logger
HH376-Series
From
$
582.44
Models In Stock

Product Overview
- Additional Features 1 MIN/MAX/AVG/REL/HOLD Functions , Auto Power-Off
- Battery life 100 hr
- Cable Length 1500 mm
- Data Interface USB, RS232
- Display Type LCD
- Display Type Note 5-digit LCD
- Height 36 mm
- Length 185 mm
- Measurement Type Temperature
- Number of Measurements 1
- Operating Temperature, Max 50 °C
- Operating Temperature, Min 0 °C
- Probe Diameter 5 mm
- Probe Length 230 mm
- Relative Humidity, Max 0.8
- Relative Humidity, Min 0%
- Samples,Max 16000
- Sampling Rate 2 Hz (2 updates/second)
- Sampling Rate Note 2 times per second
- Storage Humidity, Max 0.8
- Storage Humidity, Min 0
- Storage Temperature, Max 60 °C
- Storage Temperature, Min -10 °C
- Storage Temperature Note -10 to 60°C(14 to 140°F); 0 to 80% RH; altitude up to 2000 meters
- Supply Power Type AC, Battery
- Temperature Accuracy ± 0.1 °C
- Temperature Accuracy Note 100 to 400°C ± (0.05% reading + 0.1°C) [-148 to 752°F ± (0.05% reading + 0.2°F)]
- Temperature Measurement, Max 400 °C
- Temperature Measurement, Min -100 °C
- Temperature Resolution 0.01 °C
- Temperature Sensor Type 100 Ω RTD
- Temperature Sensor Type Note Pt100 (4-wire)
- Weight 360 g
- Width 65 mm
The HH376 is a high accuracy RTD thermometer plus a powerful data logger. The unit can store up to 16,000 records at programmed intervals. Windows® based software and USB interface cable are included to allow real time or recorded data to be read by a PC through the USB or RS232 port (RS232 cable is optional). HH376 is supplied with a precision 4-wire RTD probe with automatic locked push-pull connector for easy connection.
SPECIFICATIONS
Instrument
Numerical Display: 5-digit LCD
Measurement Range: -100 to 400°C (-148 to 752°F)
Resolution: 0.01°C (0.02°F)
Sensor Types: Platinum resistance temperature sensor for pt-100 4 wires, ALPHA = 0.00385
Environmental
Operating Temperature and Humidity: 0 to 50°C (32 to 122°F); 0 to 80% RH
Storage Temperature: -10 to 60°C (14 to 140°F); 0 to 80% RH; altitude up to 2000 meters
Accuracy: At (23 ± 5°C)
Range Accuracy: -100 to 400°C ± (0.05% reading + 0.1°C); -148 to 752°F ± (0.05% reading + 0.2°F)
Temperature Coefficient: For ambient temperatures from 0 to 18°C and 28 to 50°C, for each °C ambient below 18°C or above 28°C add the following tolerance into the accuracy spec. 0.005% of reading + 0.01°C (0.005% of reading + 0.02°F)
Sample Rate: 2 times per second
Dimensions: 185 L x 65 W x 36 mm D (meter only)
Weight: 360 g approx.
Power Requirement: 9V battery (included)
AC Adaptor: 9 Vdc (7 to 10V max)/20 mA
min (plug diameter: 3.5 x 1.35 mm)
Battery Life: Approx. 100 hrs
Probe
Sensor Type: Pt100 (4 wire)
Accuracy: Class A13, temperature probe
Measurement Range: -100 to 400°C (-148 to 752°F)
Sheath Diameter: 5.0 mm (0.2')
Sheath Length: 230 mm (9.05')
Cable Length: 1500 mm (59.0')
Read More
SPECIFICATIONS
Instrument
Numerical Display: 5-digit LCD
Measurement Range: -100 to 400°C (-148 to 752°F)
Resolution: 0.01°C (0.02°F)
Sensor Types: Platinum resistance temperature sensor for pt-100 4 wires, ALPHA = 0.00385
Environmental
Operating Temperature and Humidity: 0 to 50°C (32 to 122°F); 0 to 80% RH
Storage Temperature: -10 to 60°C (14 to 140°F); 0 to 80% RH; altitude up to 2000 meters
Accuracy: At (23 ± 5°C)
Range Accuracy: -100 to 400°C ± (0.05% reading + 0.1°C); -148 to 752°F ± (0.05% reading + 0.2°F)
Temperature Coefficient: For ambient temperatures from 0 to 18°C and 28 to 50°C, for each °C ambient below 18°C or above 28°C add the following tolerance into the accuracy spec. 0.005% of reading + 0.01°C (0.005% of reading + 0.02°F)
Sample Rate: 2 times per second
Dimensions: 185 L x 65 W x 36 mm D (meter only)
Weight: 360 g approx.
Power Requirement: 9V battery (included)
AC Adaptor: 9 Vdc (7 to 10V max)/20 mA
min (plug diameter: 3.5 x 1.35 mm)
Battery Life: Approx. 100 hrs
Probe
Sensor Type: Pt100 (4 wire)
Accuracy: Class A13, temperature probe
Measurement Range: -100 to 400°C (-148 to 752°F)
Sheath Diameter: 5.0 mm (0.2')
Sheath Length: 230 mm (9.05')
Cable Length: 1500 mm (59.0')