CT6100-Series
1, 2 or 3 Channel Programmable Process Input Circular Recorders
- Accepts Thermocouple, RTD, Vdc, and mA DC Inputs
- High Accuracy and Stability
- User Configurable via Front-Panel Keypad
- Easy-to-Use Menu-Driven Interface for Rapid Configuration and Calibration
CT6100-Series
From
$
3,712.80

Product Overview
- Accuracy ± 0.2 %
- Accuracy Detail Thermocouple: ±0.25%, RTD, Low Range (-200 to 200°C): ±0.2°C, RTD, High Range (200 to 850°C): ±0.8°C, Voltage: ±0.2%, Current: ±0.2% Linearization: "T/C Types J, K, T, N, E: ±0.1°C, -50 to 200°C, ±1°C maximum, T/C Types R and S: ±0.2°C, -50 to 200°C, ±1°C maximum, T/C Type B: ±1°C maximum RTD: ±0.1°C, -200 to 850°C"
- Additional Features 1 Transmitter Power Supply
- Approvals CE
- Chart Diameter 244 mm
- Chart Note Chart Drive: DC Stepper Motor
- Chart Speed 1 to 24 hours in steps of one hour, 2 to 31 days in steps of 1 day
- Current Measurement Range ± 20 mA
- Cutout Dimensions 288 x 356 mm
- Data protection Access by password, Internal hardware jumper lock, Lockable Door
- Deadband 0 to 10%
- Display Note 2 line x 20 character dot matrix liquid crystal with backlight and automatic temperature compensation; 9.6 mm (0.37") character height. Alarm Display: Relay status shown by red and green front panel LEDs
- Display Type LCD
- Enclosure Ratings IP55, NEMA Type 12
- Height 396 mm
- Input Note Input Isolation: Optoelectronic on 2 and 3 input models 500 Vrms channel to-channel, 500 Vrms channel-to-ground
- Input Protection ±50 Vdc; Snubber Network: 22 nF and 100 Ω across each contact
- Input Resistance Thermocouple: 10 MΩ;±2 mA: 200 Ω;±20 mA: 20 Ω;±2V, ±20V: >1 MΩ
- Length 171 mm
- Materials Steel, GF Polyester, Acrylic
- Measurement Range Note Thermocouples to BS4937 (type B minimum temp 200°C) Current: ±2 mA, ±20 mA Voltage: ±2V, +20 Vdc
- Measurement Type Thermocouple, RTD, Current, Voltage
- Mounting Panel Mounting or Surface Mounting
- Operating Temperature, Max 55 °C
- Operating Temperature, Min 0 °C
- Pen Type Pen Lift
- Pen Type Note Powered, activated from front panel; chart fast time advance possible with pens raised
- Power Consumption max 25 W
- Relative Humidity 0 to 90% RH, non-condensing
- Resolution Positioning Resolution: >0.1%
- Response Time 4.5 sec
- RTD Type 100 Ω Pt sensor; α= 0.00385
- Sensor Break Detection Thermocouple Burn-Out: Pull-Up or Pull-Down, Selectable; Sensor Fault Action: Software selectable from: high alarm/low alarm/deviation alarm/control relay (high)/control relay (low); relays de-energize in alarm state; assignable to any channel
- Sensor/Transmitter Power 12/24 Vdc @ 25 mA
- Span, Min 5°C thermocouples K, J, T, E, N and RTD; 50°C other thermocouple types, 50 mV, 200 mA
- Standard Limits of Error Error Limit: Cold-Junction Compensation Error: ±0.02°C/°C; Calibration Shift: ±10°C
- Supply Power Type AC
- Switch Rating 6 A @ 30 Vdc/250 Vac
- Switch Rating Note Type: SPDT; 30 Vdc or 250 Vac @ 6 A maximum; 150 W DC, 1660 Vac non-inductive
- Switch Type SPDT NO/NC, Selectable
- Temperature Measurement Range To BS 4937 - International thermocouple reference tables
- Temperature Stability ±0.02% FS/°C
- Thermocouple Type J, K, T, R, S, N, E, B
- Voltage Measurement Range -2 to 20 Vdc
- Width 336 mm
- Wire Code Option 3- Wire
The CT6100 Series microprocessor based circular chart recorders are available in 1-, 2-, or 3-channel variants. They combine the simplicity and clarity of pen drawing with the versatility of microprocessor control. Each channel is compatible with all industry-standard sensors and signals, including thermocouple types J, K, T, E, N, R, S, and B; Pt100 platinum RTDs; and 4 or 20 mA current loops.
Low and high measurement ranges are provided for each input type; separate range cards are not required. Multi-input versions feature optoelectronic isolation of the input stages to eliminate troublesome installation ground loops. Multislope integrating 16-bit A/D converters ensure precise measurement by sampling the input every 0.5 seconds.
Thermocouple and RTD characteristics are fully linearized. The recorders use automatic cold-junction compensation for thermocouple measurement.
All CT6100 Series models feature alarm relays. Single-pen recorders are equipped with 3 fail-safe singlepole changeover relays; 6 relays are standard on the 2- and 3-pen versions. Color-coded LEDs indicate the status of each relay. All relay functions are user selectable: setpoint values and hysteresis levels are entered directly via the keypad, while relay action and channel assignment are selected from user-friendly menus. Circuit precision is matched by the backlash-free pen-drive mechanism, which has a positioning resolution of better than 0.1%. An integral feedback potentiometer enables closed-loop monitoring of each pen position. Each recorder uses a low-maintenance stepper motor. The rotation speed of the 244 mm diameter chart is microprocessor controlled and user programmable.
All CT6100 recorders are housed in a strong molded case that can be panel or surface mounted. A gasket-lockable door protects internal components from harsh industrial environments and offers protection rated to NEMA 12 (IP55). A tough acrylic window lets the user view the chart trace, digital channel readings, and alarm status with the recorder's door closed.
Low and high measurement ranges are provided for each input type; separate range cards are not required. Multi-input versions feature optoelectronic isolation of the input stages to eliminate troublesome installation ground loops. Multislope integrating 16-bit A/D converters ensure precise measurement by sampling the input every 0.5 seconds.
Thermocouple and RTD characteristics are fully linearized. The recorders use automatic cold-junction compensation for thermocouple measurement.
All CT6100 Series models feature alarm relays. Single-pen recorders are equipped with 3 fail-safe singlepole changeover relays; 6 relays are standard on the 2- and 3-pen versions. Color-coded LEDs indicate the status of each relay. All relay functions are user selectable: setpoint values and hysteresis levels are entered directly via the keypad, while relay action and channel assignment are selected from user-friendly menus. Circuit precision is matched by the backlash-free pen-drive mechanism, which has a positioning resolution of better than 0.1%. An integral feedback potentiometer enables closed-loop monitoring of each pen position. Each recorder uses a low-maintenance stepper motor. The rotation speed of the 244 mm diameter chart is microprocessor controlled and user programmable.
All CT6100 recorders are housed in a strong molded case that can be panel or surface mounted. A gasket-lockable door protects internal components from harsh industrial environments and offers protection rated to NEMA 12 (IP55). A tough acrylic window lets the user view the chart trace, digital channel readings, and alarm status with the recorder's door closed.