Product Specs
<< See all Product Models| Brand |
|
| Accuracy | ± 1 °C (± 1.8 °F) |
| Accuracy Detail | ±1 °C (±1.8 °F) dew point (+20 to -60 °C / +68 to -76 °F); ±2 °C (±3.6 °F) dew point (-60 to -110 °C / -76 to -166 °F) |
| Applications | Air, Air Dryers, Chemicals, Commercial and Industrial, Compressor Control, Environmental and Weather, Flow Measurement, Fuel, Fuel and Gas, Gases, General Purpose, Natural Gas Compressors, Oil, Scientific Equipment, Temperature Maintenance |
| Calibration Note | Traceable 13-point calibration and certificate |
| Compensated Temperature, Max | 50 °C |
| Compensated Temperature, Min | -20 °C |
| Electrical Connection | Screw Terminal |
| Flow Rate Note | 1 to 5 Nl/min mounted in standard sampling block; 0 to 10 m/sec direct insertion |
| Galvanic Isolation | KFD0-CS-EX2.50P,KFD0-CS-EX1.50P,KFD2-STC4-EX1.H,MTL5541 |
| Housing Material | 316 Stainless Steel |
| Load Resistance | Max 250 Ω @ 12 V (500 Ω @ 24 V) |
| Measurement Type | Dew point, Moisture |
| Operating Pressure, Max | 52.5 MPa |
| Operating Temperature, Max | 60 °C |
| Operating Temperature, Min | -40 °C |
| Repeatability | 0.5 °C (0.9 °F) dew point |
| Response Time | 5 mins to T95 (dry to wet) |
| Sensor Filter | 80 µm SS sintered guard |
| Series ID | EPR-IS |
| Supply Current | 23 mA |
| Supply Voltage | 12 to 28 Vdc |
| Weight | 1.27 kg |