日本 +81 3 6822 0370 | 新しいウェブサイトへようこそ。問題やバグを見つけましたか?こちらからご報告ください
Series-400
HVACバランス調整およびファン試験用エアベロシティメーターキット、ケース付き。
- 빨간 숫자는 FPM 단위의 속도를 표시하며 검은 숫자는 인치 w.c. 단위의 공기 압력을 표시합니다.
- 표준 조건의 공기에 대한 변환 표가 필요 없습니다.
- 피토관 및 휴대용 케이스와 함께 제공됩니다.
製品の概要
- 공기 속도, 최소 400 FPM
- 측정 유형 공기 속도
- 작동 온도, 최대 150 °F
- 압력, 최소 0 inH₂O
- 압력 유형 해당 없음
- 증명압 최대 100 psi
- 시리즈 ID Series-400
Supported Applications
The Dwyer Series 400 Air Velocity Meter is a dual-purpose instrument designed for measuring air velocity and static pressure. Documented applications include balancing air conditioning systems, testing fan and blower discharge and inlet pressures, measuring furnace draft, and determining air filter pressure drop.
Operating Conditions & Performance
The series operates with process temperatures up to 150°F (65°C) and is rated for total pressures to 100 psig (6.89 bar). Measurement capabilities cover a velocity range of 400 to 12,600 FPM and pressure ranges including 0 to 10 inch w.c., depending on the specific model configuration.
Accuracy specifications vary by application environment:
- Laboratory applications: ±2% accuracy (requires extreme care, duct diameter of 4" or greater, and smooth straight duct sections).
- Field tests: ±5% accuracy.
The instrument uses Dwyer gage fluid only; other fluids, solvents, or cleaning agents may damage the gauge. Cleaning is restricted to mild soap and water.
Configuration Options
The Series 400 offers configurable options based on kit contents and scale requirements. The series supports combination inclined/vertical scales with specific minor divisions:
- Inclined Scale: .01 inch w.c., range 0 to 1.0.
- Vertical Scale: .10 inch w.c., range 1-10.
Kits include a manometer, Pitot tube (e.g., 18 in length for the 400-10 model), necessary tubing and connectors, red gage fluid, a slide chart, and a carrying case. The complete kit weighs only 13-1/4 lb with equipment.
Model numbering follows the format 400-<MODEL>-<KIT/GAGE>, such as 400-10-Kit or 400-10-Gage. The series includes options for extra long steel cases to accommodate longer Pitot tubes, available by contacting the factory.
Key Product Differences
Variants within the Series 400 differ primarily in kit contents and scale ranges:
- Kits vs. Gages: Models ending in "-KIT" include a Pitot tube, tubing, connectors, fluid, slide chart, and carrying case. Models ending in "-GAGE" (e.g., 400-10-Gage) contain only the manometer.
- Pitot Tube Length: The 400-10 model includes an 18-inch Pitot tube; other models may vary or exclude the probe entirely in gage-only configurations.
アクセサリー
consumables
400 - Instruction Manual - 400_IOM_pdf ユーザーマニュアル
Red Gage Fluid - French - dwyer_100av_series_manual_red_gage_fluid_fr_xzbu45qdlilq_fr_fr_pdf_pdf ユーザーマニュアル
Blue Gage Fluid - Spanish - dwyer_400_series_manual_blue_gage_fluid_sp_1ayjztlo6163d_es_es_pdf_pdf ユーザーマニュアル
Blue Gage Fluid - German - dwyer_400_series_manual_blue_gage_fluid_gr_1ex533lqsmov_de_de_pdf_pdf ユーザーマニュアル
Blue Gage Fluid - English - dwyer_400_series_manual_blue_gage_fluid_1mjwvaleeitws_en_us_pdf_pdf ユーザーマニュアル