Besoin d'aide? Appelez-nous: 1-800-663-4209
HX300A-Series
Humidity/Temperature Transmitters
- Rápido, estable y preciso
- Carcasa estanca
- Compacto, fácil de usar
- Excelente estabilidad a largo plazo
HX300A-Series
À partir de
Aperçu du produit
- Exactitud ±2,5 % HR, ±0,3 °C
- Conexión eléctrica Prensaestopas para cables de 5 a 10 mm con bloque de terminales atornillables
- Clasificaciones del alojamiento IP65
- Estabilidad a largo plazo Mejor que 1 % HR por año (típico)
- Temperatura de operación, máx. 50 °C
- Temperatura de operación mínima 0 °C
- Tiempo de respuesta 15 s
- Material del sensor Filtro sinterizado
- ID de serie HX300A-Series
- Tipo de alimentación eléctrica CC
- Tensión de alimentación 12 a 40 V CC
- Compensación de temperatura 0,008 % HR/°C (efecto @ 05 HR)
- Peso 150 g
Supported Applications
The Omega HX300A Series consists of wall-mount transmitters designed for industrial environments requiring fast, stable, and accurate measurements. The series employs a highly stable thin-film polymer capacitor to sense relative humidity and a high-accuracy, thin-film RTD (PT100 Ω) compliant with IEC 751 and DIN 43760 standards to accurately sense temperature.
Operating Conditions & Performance
The HX300A Series operates within an ambient temperature range of 0 to 50°C (32 to 122°F). Measurement capabilities include a humidity range of 0 to 100% RH @ 0 to 50°C (32 to 122°F) and a temperature measurement range of 0 to 100°C (32 to 212°F).
Performance specifications at 25°C include an accuracy of ±2.5% RH, ±0.3°C with long-term stability better than 1% RH per year (typical). The response time is <15 seconds (90% @ 25°C in moving air at 0.5 m/sec), and temperature compensation is rated at 0.008% RH/°C (effect @ 05 RH).
The unit features a sintered filter for sensor protection and utilizes an ABS plastic watertight enclosure to NEMA 4 (IP65). Electrical connections are provided via liquid-tight nylon cable bushing fitting 5 to 10 mm dia. (0.2 to 0.4") cables.
Key Specifications
- Housing: ABS plastic watertight enclosure to NEMA 4 (IP65)
- Sensors - Humidity: Thin-film capacitor
- Sensors - Temperature: RTD PT100 Ω, IEC 751, DIN 43760
- Weight: Approx 150 g (5.3 oz)
- Approvals/Certifications: CE, NIST-traceable calibration available via CAL-3-HU accessory
Configuration Options
The series supports multiple output configurations and sensor combinations to address specific monitoring needs. Power supply requirements are 12 to 40 Vdc (24 Vdc recommended).
Output Signals
- Current Output: 4 to 20 mA, 2-wire with ±15% scalable zero and span adjustment.
- Voltage Output: 0 to 1 V (primary series specification).
Sensor Configurations
- Humidity only models (HX302A family)
- Temperature and Humidity combined models (HX303A family)
Accessories & Options
The series is offered with various power supply options including:
- PSR-24S: Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, screw terminals, UL.
- PSR-24L: Regulated power supply, US plug, 90 to 264 Vac input, 24 Vdc output, 400 mA, stripped leads, UL.
- PSR-24L-230: Regulated power supply, European plug, 230 Vac input, 24 Vdc output, 400 mA, stripped leads, CE.
- PSU-93: Unregulated power supply, 16 to 23 Vdc, 300 mA max, screw terminal.
Key Product Differences
The HX300A Series is differentiated by sensor type and output signal format. The model number guide follows the pattern: HX300A - <MODEL> - <OUTPUT TYPE>.
- HX302AC: Wall mount RH transmitter, 4 to 20 mA current output.
- HX302AV: Wall mount RH transmitter, 0 to 1 voltage output.
- HX303AC: Wall mount RH/temperature transmitter, 4 to 20 mA current outputs.
- HX303AV: Wall mount RH/temperature transmitter, 0 to 1 voltage outputs.