Singapore 1800 622 1156 | Intl +65 6415 5353 | Welcome to our new site - found a bug? Please report it here
TXUN-FD-Temp-Transmitter
Field Mounted Hart Protocol Temperature Transmitter
- RTD, TC, Ω, and Bipolar mV Input and Analog Output
- High Definition Local Operator Interface (LOI) with 3 Optical Buttons
- Selectable Red or White Backlight
- Ex D Explosion Proof/Flame Proof in Aluminum or 316 Stainless Steel Version
TXUN-FD-Temp-Transmitter
From
SGD
1,680.33
Product Overview
- Accuracy ± 0.05 %
- Additional Features 1 Explosion Proof Aluminum/Stainless Steel Housing
- Approvals EU RO Mutual Recognition Type; Approval: MRA0000009; ATEX 2014/34/EU: DEKRA 15 ATEX 0058 X; IECEx: IECEx DEK 15.0039 X; FM: FM16US0009X/FM16CA0010X; CSA: 70024231 EAC Ex TR-CU 012/2011: RU C-DK.GB08.V.01316; INMETRO: DEKRA 15.0014 X; NEPSI: GYJ15.1336X, GYJ15.1337X and GYJ15.1338X; SIL: Hardware assessed for use in SIL applications Authority Requirements: EMC: 2014/30/EU, EAC: TR-CU 020/2011
- Calibration Adjustment ± 50 % Max
- Calibration Temperature, Max 28 °C
- Calibration Temperature, Min 20 °C
- Data Interface HART 7, HART 5
- Dimensions Note Aluminum: 109.3 H x 145 W x 126 D mm, Stainless: 107.4 H x 145 W x 124 D mm Dimensions: Ø 110 mm
- Display LCD
- Display Note 96 x 64 pixels Resolution, Selectable ON/OFF and Selectable white or red colour Backlight Number of Digits: 5
- Electrical Output 4 to 20 mA
- EMC Immunity Note Standard EMC Imunity: <±0.1% of span; Extended EMC Immunity: NAMUR, NE21, A Criterion, Burst: <±1% of span; Wire Size: 0.13 x 1.5 mm²
- EMC Standard EN61010-1, EN61326-1
- Enclosure Ratings IP54, IP66, IP68, NEMA Type 4X
- Height 109 mm
- Input Control Signal -800 to 800 mV
- Input Control Signal Note Min. Measurement Range (Span): 2.5 mV
- Input Resistance 10 MΩ
- Isolation Test/Working: 1.5 kVac/50Vac
- Length 104 mm
- Load Resistance ≤ (Vsupply - 12) / 0.023 Ω
- Long Term Stability ±0.1% of span/year
- Operating Temperature, Max 85 °C
- Operating Temperature, Min -40 °C
- Operating Temperature Note -40 to 85°C (with Silicone O-ring),-20 to 85°C (with FKM Oring)
- Output Detail Signal Range: 4 to 20 mA,Min. Signal Range: 6 mA,NAMUR NE43 Upscale/Downscale: 23 mA/3.5 mA
- Output Load ≤ (Vsupply - 10) / 0.023 Ω
- Output Signal Current
- Potentiometer Input Range 0 to 7000 Ω
- Programmable Functions HART Setup
- Relative Humidity 0 to 100% RH Condensing
- Resolution Input Signal Dynamics: 22 bit; Output Signal Dynamics: 16 bit
- Response Time 1 to 60 sec; Update Time : 440 msec
- RTD Input Note Cable Resistance, Per Wire: 50 Ω; Sensor Current: 0.2 mA
- RTD Type 50/100/200/250/500/1000 Ω Pt Sensor, 50/100/120/1000 Ω Ni Sensor
- Screw Terminal Torque 0.4 Nm
- Sensor Break Detection Programmable 3.5 to 23 mA
- Series ID TXUN-FD-Temp-Transmitter
- Signal/Noise Ratio > 60 dB
- Start Up Time 5 Sec
- Storage Temperature, Max 85 °C
- Storage Temperature, Min -40 °C
- Supply Power Note Intrinsically Safe: 10 (12 - with backlight)...30 Vdc, Other: 10 (12 - with backlight)...35 Vdc
- Supply Power Type DC
- Supply Voltage 10 to 30 Vdc
- Temperature Measurement Range J: -100 to 1200°C, K: -180 to 1370 °C, T: -200 to 400 °C, E: -100 to 1000 °C, R: -50 to 1760 °C, S: -50 to 1760 °C, N: -180 to 1300 °C, B: 0 to 1820 °C, W3: 0 to 2300 °C, W5: 0 to 2300 °C, L: -200 to 900 °C, LR: -200 to 800 °C, RTD: -200 to 850 °C
- Thermocouple Type J, K, T, E, N, R, S, U, B, C, D, L, LR
- Vibration ±4g
- Vibration Note EC 60068-2-6; 2 to 25 Hz: ±1.6 mm; 25 to 100 Hz: ±4 g
- Weight 1.3 kg
- Weight Note Aluminum: 1.3, Stainless: 2.8
- Width 23.5 mm
- Wire Type Stranded
Supported Applications
The Omega TXUN-FD is a field-mounted HART temperature transmitter designed for RTD, TC (thermocouple), Ohm, and bipolar mV inputs. The series supports general purpose installations as well as hazardous area applications with Ex d explosion-proof / flameproof ratings in aluminum or 316 stainless steel housings.
Operating Conditions & Performance
The transmitter operates within an ambient temperature range of -40°C to +85°C when equipped with a silicone O-ring, and -20°C to +85°C with an FKM rubber O-ring. Storage temperature is rated from -40°C to +85°C. The device functions in relative humidity conditions of 0...100% RH (condensing). Environmental protection ratings include IP54 / IP66 / IP68 and type 4X.
Performance specifications feature a display resolution of 96 x 64 pixels with a response time from button to display of < 150 ms. The mechanical design accommodates vibration per IEC 60068-2-6, specifically ±1.6 mm at 2...25 Hz and ±4 g at 25...100 Hz.
Power Supply Requirements
- 10...30 VDC (standard)
- 12...30 VDC with backlight enabled
- 10...35 VDC (extended range option)
- 12...35 VDC with backlight enabled
Configuration Options
The TXUN-FD series is configurable across housing materials, local operator interface options, conduit threads, and approval levels. The model number guide follows the structure: TXUN-FD - <Housing> - <Local operator interface> - <O-ring> - <Conduit thread> - <Paint type> - <Transmitter> - <Approvals> - <Cover color>.
Housing and Materials
- Low copper aluminum (AL): Available with Epoxy or Epoxy + polyurethane paint.
- 316 Stainless steel (RF): Paint type is None; suitable for corrosive environments.
Local Operator Interface and Display
- Configurable with or without a local operator interface (LOI).
- The LOI features 5 digits, selectable white or red backlight, and error indication in the same colors.
- Cover color options include Red, Gray, or Steel finish depending on housing type.
Connections and Seals
- O-ring materials: Silicone rubber (-40 to +85°C) or FKM rubber (-20 to +85°C).
- Conduit threads (D1, D2 & D3): M20x1.5 6H or ½ NPT mod.
- Electrical connection: Accommodates wire sizes of 0.13...1.5 mm² / AWG 26...16 stranded wire with a screw terminal torque of 0.4 Nm.
Approvals and Certifications
The series supports the following approvals: UL LISTED, CE, EAC, Ex, CSA, IEC, IECEx, INMETRO, and NCC. Hazardous area configurations are available alongside general purpose models.
Key Product Differences
Variants differ primarily by housing material (aluminum vs. 316 stainless steel), the presence of a local operator interface, O-ring type determining temperature range, and conduit thread style. The aluminum version is offered with red or gray cover colors, while the stainless steel version features a steel finish.
Models without a transmitter unit are supplied as connection kits (indicated by "No" in the Transmitter column of the order guide). Accessories include M20 plugs for various enclosure and O-ring combinations (part numbers 8550 series), NPT plugs, pipe-mounting brackets (P5-B-N), display spare parts with or without LOI, and a 5-point calibration certificate.