  • Fast 1-second reading rate captures rapid environmental changes
  • Multiple start/stop function supports repeated test cycles or studies
  • Stores up to 1,000,000 readings; easy data download via USB with MadgeTech 4
  • Measures and logs humidity and temperature with ±3%RH, ±0.5 °C accuracy
OM-CP-RHTEMP101A-Series
From  203,42
Humidity and Temperature Data Logger, Part of the NOMAD® Family OM-CP-RHTEMP101A-Series OM-CP-RHTEMP101A-Series
loading all models
Product Overview
  • Battery life up to 10 yr
  • Battery Size and Type 3.6V Li-ion
  • Calibration Digital calibration through software
  • Data Interface USB
  • Data Interface Note USB(interface cable required);115200 baud
  • Form factor Benchtop
  • Height 16 mm
  • LED Indicators Green LED Blinks: 10 second rate to indicate logging 15 second rate to indicate delay start mode ;Red LED Blinks: 10 second rate to indicate low battery and/or full memory 1 second rate to indicate an alarm condition
  • Length 36 mm
  • Memory 1000000
  • Memory Modes Yes
  • Memory Note 1,000,000 readings per channel;software configurable memory wrap; 500,000 readings in multiple start/stop mode
  • Number of Measurements 1
  • Operating Humidity 0 to 95% RH
  • Operating Temperature, Max 80 °C
  • Operating Temperature, Min -40 °C
  • Process Temperature, Max 80 °C
  • Process Temperature, Min -40 °C
  • Real Time Recording The device may be used with PC to monitor and record data in real-time.
  • Relative Humidity Accuracy ± 3% RH
  • Relative Humidity Accuracy Note 10 to 40ºC (50 to 104ºF); 10 to 80% RH
  • Relative Humidity, Max 0.95
  • Relative Humidity, Min 0%
  • Relative Humidity Resolution 0.001
  • Relative Humidity Sensor Type Semiconductor
  • Samples,Max 500000
  • Sampling Rate 1 Hz (1 update/second)
  • Sampling Rate Note 1 reading every second to 1 every 24 hours.
  • Start Modes Adjustable Delay Start, Immediate Start, Push-button
  • Stop Modes Adjustable Delay Stop, Manual Through Software
  • Supply Power Type Battery
  • Temperature Calibration Accuracy ± 0.5 °C
  • Temperature Resolution 0.01 °C
  • Temperature Sensor Type RTD
  • Temperature Sensor Type Note Precision RTD Element
  • Time Accuracy ± 1 minute/month
  • Weight 24 g
  • Width 56 mm

Description

The OM-CP-RHTEMP101A is part of a new series of low cost, state-ofthe-art data logging devices. The OM-CP-RHTEMP101A offers a 10 year battery life, a 1 second reading rate, a multiple start/stop function, ultra-high speed download capability, 500,000 reading storage capacity, optional memory wrap, precision RTD sensing element, battery life indicator, optional password protection, programmable high-low alarms and more.

Using the software, starting, stopping and downloading from the OM-CP-RHTEMP101A is simple and easy. Graphical, tabular and summary data is provided for analysis and data can be viewed in °C, °F, K,
°R, %RH, mg/ml water vapor concentration and dew point. The data can also be automatically exported to Excel® for further calculations.

The OM-CP-RHTEMP101A major leap forward in both size and performance. Its real time clock ensures that all data is time and date stamped. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged. Its small size allows it to fit almost anywhere. Data retrieval is simple. Plug it into an available USB port and the easy to use Windows software does the rest. The software converts your PC into a real time strip chart recorder. Data can be printed in tabular format and can also be exported to a text or Microsoft Excel file.

The OM-CP-RHTEMP101A designed with our customers in mind. There are free firmware upgrades for the life of the product so that data loggers already deployed in the field can grow with new technological developments. Units do not need to be returned to the factory for upgrades. The user can do this automatically from any PC.