India +91 8925807296 | Welcome to our new website, found an issue or bug? Please report it here
RHCM-40
Optical Chilled Mirror Dew Point Transmitter
- Wall, Duct and Pipe Mount Models
- Remote Type Models
- 4-Wire RTD Measurement for Dew Point
- ± 0.2°C Accuracy for Both Dew Point and Temperature
Product Overview
- Accuracy ± 0.2 °C (± 0.36 °F)
- Additional Features 1 For Dew Point - 2 stages of cooling (TEC) with a depression of 60°C; Standard Chrome Plated Mirror
- Cable Length 3 m
- Cable Length, Max 15.2 m
- Data Interface RS232
- Dew Point Temperature, Max 60 °C
- Dew Point Temperature, Min -40 °C
- Electrical Output 4 to 20 mA, 0 to 5 Vdc, 0 to 10 Vdc
- Height 130 mm
- Hysteresis Negligible
- Length 76 mm
- Linear Flow Rate, Max 3000 FPM
- Current, Max 1 A
- Output Detail 4 to 20 mA: 0 to 250 Ω 0 to 5/10 Vdc: 1 kΩ or higher
- Pressure Limits Up to 100 psi
- Measurement Pressure, Min 0 psi
- Pressure Note 100 psig maximum (higher pressure available on special order)
- Relative Humidity 1 to 95%
- Repeatability ± 0.05°C
- Response Time 1°C/second not including settling time
- Series ID RHCM-40
- Supply Power Type DC
- Supply Voltage 24 Vdc
- Temperature Measurement, Max 60 °C
- Temperature Measurement, Min -40 °C
- Weight 1.4 kg
- Width 130 mm
Supported Applications
The Omega RHCM-40 Series is an optical chilled mirror dew point transmitter designed for continuous moisture content measurement in gases. The instrument utilizes the chilled mirror dew point condensation principle and a platinum resistance RTD for temperature measurement. Documented applications include process control and energy management, with optional capabilities for air temperature and relative humidity monitoring.
Operating Conditions & Performance
The RHCM-40 series provides specific performance metrics based on the chilled mirror sensor technology:
- Dew Point Range: -40 to 60°C (-40 to 140°F) (Standard)
- Air Temperature Range: -40 to 60°C (-40 to 140°F) (Optional)
- Relative Humidity Range: 1 to 95% RH (Optional)
- Accuracy: ±0.2°C for both dew point and temperature
- Repeatability: ±0.05°C
- Hysteresis: Negligible
- Response Time: 1°C/second not including settling time
- Pressure Rating: 100 psig maximum (higher pressure available on special order)
The instrument is powered by a standard 24 Vdc ±20% supply with a 1 A maximum, or an optional AC Universal Supply of 85 to 230 Vac.
Configuration Options
The series supports multiple mounting configurations and enclosure materials tailored for different installation environments:
- Housing Materials: IP65 plastic enclosure (for -PL models) or Aluminum enclosure (for -AL models)
- Mounting Options: Wall Mount, Duct Mount, Pipe Mount, and Remote Mount
- Process Connections: Includes 1¼ NPT brass pipe fitting for pipe mount configurations; duct mount flange kits are available for specific models.
- Cable Lengths (Remote): Standard remote cable is 3 m (10') with a maximum of 15.2 m (50'); Remote Mount -AL-R includes a 3 m (10') cable between sensor and electronics, while -PL-R models include a 1.8 m (6') cable.
- Outputs: Configurable for 4 to 20 mA (load resistance: 0 to 250 Ω), 0 to 5 Vdc or 0 to 10 Vdc (load resistance: 1 kΩ or higher), RS232 bi-directional serial output, and two programmable alarm relays.
- Relay Specifications: Contact rating is Normally open (NO) 3 A at 25 Vac, or 30 Vdc. Relays are assignable to Clean mirror/Alarm 1 (K1) or Servolock/Alarm 2 (K2).
Sensor and Mirror Options: The standard mirror is chrome plated; optional mirrors include gold or platinum coatings. Wetted materials correspond to the selected mirror finish.
Optional Features & Accessories
- Air Temperature Sensors: Available as Remote mount (-ATDT/R), Wall mount (-ATDT/W for -W models only), or Integrated 100Ω Pt RTD sensor embedded in the dew point sensor (-ATDT/I).
- Display: Two Line LCD (Graphical Display with 3 Parameters Displayed at One Time) available as option -RD for -PL models.
- Sample Chambers: Flow-Through Sample Chamber options include low pressure sample chamber (-SC), max pressure 100 psig flow through (-SC1), and High Pressure (20 bar).
- Power & Control: Universal power supply option (-ACDT) for 85 to 230 Vac, 50/60 Hz; Sample module universal operation (-SMU); Aspirator kit (-AR).
- Cabling Extensions: Additional cable length up to 15.2 m (50') via option -ECDT.
Key Product Differences
The RHCM-40 series is distinguished by enclosure material and mounting style, defined in the model number guide as RHCM-40 - <MOUNTING> - <ENCLOSURE> - <OPTIONAL FEATURES>.
- Enclosure Types: Models with suffixes starting with "PL" feature an IP65 plastic enclosure. Models with suffixes starting with "AL" utilize an aluminum enclosure.
- Mounting Configurations:
- -W: Wall mount (includes 2-staged chilled mirror sensor, analog outputs, RS232, alarm relays)
- -D: Duct mount (includes duct mount flange kit)
- -R: Remote mount (separates sensor and electronics with a cable; -PL-R includes 1.8 m (6') cable, -AL-R includes 3 m (10') cable)
- -P: Pipe mount (includes 1¼ NPT brass pipe fitting rated for 100 psi maximum)
Note regarding the LCD option (-RD): When ordered, all three parameters are displayed on the screen; however, only two of these parameters can be assigned to the two analog outputs.
PDFs & Manuals