Skip to search Skip to main content Skip to navigation

India +91 8925807296 | Welcome to our new website, found an issue or bug? Please report it here

We noticed you are in , please visit your local site by clicking the below link for regional offers and support.

MDSi8-MDSSi8

Benchtop Digital iSeries Thermometers Ten-Channel Models

  • Portable, Rugged Metal Benchtop Enclosure with Tilt Handle
  • 10-Channel Models
  • Built Around OMEGA’s Award Winning iSeries Meters
  • 5-Year Warranty
MDSi8-MDSSi8
From $ 1,854.30
Benchtop Digital iSeries Thermometers Ten-Channel Models MDSi8-MDSSi8 MDSi8-MDSSi8
loading all models
Product Overview
  • A/D Conversion Dual slope
  • Accuracy ± 0.4 °C
  • Accuracy Detail RTD (Pt, 0.00385, 100, 500, 1000 Ω) : 0.4°C, RTD (Pt, 0.00392, 100, 500, 1000 Ω) : 0.4°C
  • Additional Features 1 Benchtop Enclosure
  • Common Mode Rejection 120 dB
  • Control Output 2 SPDT Relay
  • Current Measurement Range 0 to 20 mA
  • Data Interface RS232, RS422, RS485
  • Data Interface Note RS232/RS422/RS485: Selectable from menu; both ASCII and MODBUS protocol selectable from menu; programmable 300 to 19.2 Kb; complete programmable setup capability; program to transmit current display, alarm status, min/max, actual measured input value and status Selectable from menu; both ASCII and MODBUS protocol selectable from menu; programmable 300 to 19.2 Kb; complete programmable setup capability; program to transmit current display, alarm status, min/max, actual measured input value and status RS485: Addressable from 0 to 199
  • Decimal Selection 0, 0,001, 0,01, 0,1
  • Digital Filter Programmable
  • Display Color Red, Green, Amber
  • Display Size 10.2 mm (0.40 in)
  • Display Type 4 Digit, 9 Segment LED
  • Display Type Note 4-digit, 9-segment LED, 10.2 mm (0.40") i32, i16, i16D, i8DV: 21 mm (0.83") i8: 10.2 mm (0.40") and 21 mm (0.83"), i8DH—GREEN, AMBER, and RED programmable colors for process variable, setpoint and temperatureunits
  • Electrical Connection Sub-D Female Miniature Connector
  • Electrical Connection Note DB-9 connector on rear panel; optional alarm 1 and 2 (programmable)
  • Form factor 1/8 DIN
  • Frequency 50 to 60 Hz
  • Impedance Input 10 MΩ for 100 mV, 1 MΩ for 1 or 10 Vdc
  • Input Configuration Single-ended
  • Normal Mode Rejection 60 dB
  • Number of Inputs 1
  • Offset Adjustment -1999 to +9999
  • Output Detail Optional Alarm 1 and 2 (Programmable) Type: Form “C” SPDT relays Operation: High/low, above/below, (limited output to 30 Vrms only) band, latch/unlatch, normally open/normally closed and process/deviation; front panel configurations Optional Analog Output (Programmable): Non-isolated, retransmission 0 to 10 Vdc or 0 to 20 mA, 500 Ω max (output 1 only). Accuracy is ± 1% of FS when following conditions are satisfied. 1) Input is not scaled below 1% of input FS 2) Analog output is not scaled below 3% of output FS ;
  • Polarity Unipolar
  • Reading Rate 3 samples/sec
  • Resolution 1°/0.1°; 10 µV process
  • Series ID MDSi8-MDSSi8
  • Set Point Adjustment -1999 to 9999 counts
  • Span Adjustment 0.001 to 9999 counts, programmable
  • Step Response 0.7 sec for 99.9%
  • Supply Voltage 90 to 240 Vac
  • Temperature Measurement Range RTD (Pt, 0.00385, 100, 500, 1000 Ω) : -200 to 900°C, RTD (Pt, 0.00392, 100, 500, 1000 Ω) : -200 to 850°C
  • Temperature Stability 0.04°C/°C

Supported Applications

The Omega MDSi8 and MDSSi8 series are benchtop digital thermometers designed for single-channel (10-point) or ten-channel monitoring. These universal instruments handle temperature inputs from common thermocouple types, multiple RTD configurations, as well as process voltage and current signals. The ¼ DIN size meters feature a rugged metal enclosure suitable for laboratory benches or industrial control panels where large-format display visibility is required.

Operating Conditions & Performance

The series supports measurement ranges spanning from -50 to 40°C/40 to 1768°C up to high-temperature thermocouple limits such as 0 to 2320°C/32 to 4208°F for Type B inputs. Accuracy specifications vary by input type: process voltage and current ranges offer ±0.03% reading accuracy, while temperature measurements range from ±0.4°C (±0.7°F) for Types J, L, C, RTD, and N, to ±1.0°C/±1.8°F or ±1.0°C/±0.5°C depending on the thermocouple type.

Resolution is 1° or 0.1° for temperature inputs and 10 μV for process signals. The response time is specified as 0.7 s for 99.9% of a step change. Power requirements are universal at 90 to 240 Vac, 50 to 60 Hz.

Configuration Options

The series offers configurable models based on channel count and input type:

  • MDSi8 / MDSi8A: Single-channel benchtop thermometers with universal inputs (thermocouple, RTD, process voltage/current). The 'A' suffix indicates inclusion of an analog output.
  • MDSi8S: A dedicated single-channel strain meter variant.
  • MDSSi8-TC / MDSSi8A-TC: Ten-channel models with dedicated thermocouple inputs; 'A' adds analog output capability.
  • MDSSi8-RTD / MDSSi8A-RTD: Ten-channel models configured for 100 Ω RTD inputs, with optional configuration for 500 or 1000 Ω RTDs on ten-channel units only.
  • MDSSi8-PV / MDSSi8A-PV: Ten-channel process meters dedicated to voltage and current inputs.

Communication and control options include isolated RS232/RS485-422 serial ports, Ethernet with an embedded web server (not available on 'A' models), and dual alarm relays rated at 3 A @ 120/240 Vac. Analog outputs provide 0 to 10 Vdc or 0 to 20 mA signals.

Key Product Differences

The primary distinction between the MDSi8 and MDSSi8 families is channel capacity: single-channel versus ten-channel dedicated inputs. Within each family, models are differentiated by input type (universal vs. thermocouple/RTD/process) and output features.

  • Analog Output: Models with the 'A' suffix include an isolated programmable analog output; these units do not support Ethernet options.
  • Communication Options: The '-C24' option provides RS232/RS485-422, while '-EIT' adds Ethernet connectivity. A combined '-C4EIT' option is available for both serial and network communication on compatible models.
  • Alarm Relays: The '-AL' suffix denotes dual form 'C' SPDT alarm relays.

The series includes a 5-year warranty and carries UL, cUL, and RoHS 2 Compliant certifications. Power cords are available for North America (120 Vac), Continental Europe, UK, Italy/Ireland, Denmark, or stripped ends for global installation up to 250 Vac.