Model-670
Fume Hood Monitor with Hot Wire Sensor & Mounting Kit for Safe Ventilation
- Flexible surface or flush mounting
- LED safe and alarm status indicators
- Audible alarm
- Sash alarm input
- Night set-back input
Model-670
From
AUD
1,793.03
Product Overview
- Accuracy ± 10 %
- Alarm Indication Red LED and Audible Alarm
- Control Output Relay
- Horn Silence Yes-Temporary and Permanent
- LED Indicators Red: Alarm; Green: Normal
- Low Air Velocity Alarm Delay 5 s
- Mounting Semi Flush, Flush or Surface Mounted
- Operating Temperature, Max 86 °F
- Operating Temperature, Min 55 °F
- Relay Input For Night Setback 2-Wire rated for 24 Vdc usage
- Relay Output 5 A @ 250 Vac
- Sash High Indication 2-Wire Micro Switch or 3-Wire Proximity Switch Input, rated for 24 Vdc usage
- Series ID Model-670
- Storage Temperature, Max 150 °F
- Storage Temperature, Min -40 °F
- Supply Power Note 15 Vdc 500 mA; 120 Vac, 60 Hz power transformer included
- Supply Power Type AC/DC
- Weight 5 oz
Supported Applications
The Dwyer Model 670 Fume Hood Monitor is designed for fume hood face velocity air flow measurement in laboratory environments. It continuously senses airflow through the face of a fume hood to ensure safe levels of fresh air are exhausting, utilizing a hot wire sensor capable of detecting very low flows common on these units.
Operating Conditions & Performance
The series measures air velocity within an alarm range of 30-400 FPM (0.15-2.0 m/s) with an accuracy of face velocity ±10%. A fixed low air velocity alarm delay is set to 5 s. The unit operates in ambient temperatures from 55 to 86°F (13 to 30°C), while storage temperature limits are -40 to 150°F (-40 to 65°C). Power requirements include 15 VDC at 500 mA, with a 120 VAC, 60 Hz power transformer included.
Configuration Options
The Model 670 provides visual and audible status indication alongside configurable inputs for facility integration. Key capabilities include:
- Status Indication: Visual LEDs (Red: Alarm, Green: Normal) and an audible alarm with temporary or permanent horn silence options.
- Relay Output: Low air flow alarm relay rated for 5 A @ 250 VAC.
- Inputs: Relay input for night setback (2-wire, rated for 24 VDC usage) and sash high indication using a 2-wire micro switch or 3-wire proximity switch input (rated for 24 VDC usage).
- Communication: RS232 comm port connectable via serial interface to a LAN network.
- Mounting & Installation: Semi-flush, flush, or surface mounted using the included bracket. The unit ships with 24" of tubing for connecting to the inside of the hood wall.
The sensor should be positioned to see room pressure; ideal placement is 100mm back from sash glass and 100mm higher than normal sash opening height. Tube lengths exceeding 1 meter or smaller diameters may restrict airflow, causing instability or incorrect readings.
Key Product Differences
The series consists of the Model 670 Fume Hood Monitor configuration. The unit weighs 5.0 oz (141 g) and carries CE agency approval. Dimensions are listed as 5-45/64 [145.03] inches in height, 3-29/64 [87.88] inches in width, with a depth of 1-1/16 [27.10].