hi how i can connect it to o2 generator @65 psi thank you
Good Afternoon,
Intended to read Air flow, does the display can be selected to show the flow in PPH?
Date published: 2021-05-06
If the meter recorded a flow when set to the incorrect media, is there a trusted conversion to post-process data?
Thank you for your inquiry. Unfortunately, no.
Date published: 2025-06-03
What are the voltage levels of the RS232 signals? -12V to +12V?
Thank you for your inquiry. The mass flow meters have by factory default a 0 to 5V signal output or as an option a 4 to 20 mA and the RS232 is available as serial communication.
Date published: 2022-10-19
Hi, is FMA-1601A usable for LPG with H2S and RSH mixture?
Thank you for your inquiry about the mass flow meter. There is an option listed in themanual for the '-LSS' version that handles corrosive gases such as the H2S listed. You may submit a request for a quote to flow@omega.com with the estmated flow range to prepare the part number required.
Date published: 2022-12-03
What is the minimum inlet pressure of the 0-5 SLPM model?
Thank you for your inquiry. The minimum pressure required is 1 PSI
Date published: 2024-08-07
Can this sensor measure the flow rate of a duct with a mixture of air, dust, and particles? Is this sensor compatible with the program NodeRED?
Thank you for your inquiry. It is not recommended to have suspended particles in the air stream and will result in clogging the internal flow path and cause damage. This flow meter requires that the air flow be clean and dry and if the NodeRED can accept an analog input from the flow meter's output, yes.
Date published: 2024-12-09
I am interested in using Modbus protocol via RS232/RS485 to poll data from this instrument. Is this possible? And if so, what register address corresponds to what type of data (ie Mass flow, Volumetric flow, Pressure and Temperature)?
Thank you for your inquiry. The manual for the MODBUS communication protocol is found in the Downloads & Software Manuals Tab.
Date published: 2024-02-11
Hi, does this flow meter work for steam at low vapor pressure (< 1 psia)?
Thank you for your inquiry. Unfortunately, the mass flow meter is best for only clean, dry gases, not for steam applications.
Date published: 2025-02-14