FMA-LP2600A-Series
Contrôleurs de débit massique de gaz à faible chute de pression pour gaz propres
Modèles en stock
- 130+ Gas Calibrations Including Pure and Mixed Gases
- Pressure, Temperature, and Mass Flow Simultaneously Displayed
- Easy-to-Use Pushbutton Interface
- NIST Traceability Standard
FMA-LP2600A-Series
Depuis
C$
3,653.53
Modèles en stock

L'aperçu du produit
- Précision ± 0.8 %
- Détail de la précision ± (0.8% of reading + 0.2% FS)
- Additional Features 1 Power Supply Included, Volumetric Flow Rate as Standard Output
- Calibration NIST Certificate
- Data Interface RS232
- Display Type LCD
- Electrical Connection 8-pin mini DIN
- Electrical Output 4 to 20 mA
- Excess Flow Rate 2.4% FS
- Height 3.9 in
- Humidity Range 0 to 100% RH non-condensing
- Input Control Signal 0 to 5 Vdc
- Media Compatibility Air, Gases, Nitrogen, Oxygen
- Operating Pressure, Max 50 psi
- Operating Pressure, Min 0 psi
- Operating Temperature, Max 50 °C
- Operating Temperature, Min -10 °C
- Pressure Drop, Max 0.06 psi
- Pressure Drop Note 0.06 Pressure Drop at FS psid venting to 1 ATM
- Repeatability ± 0.2%
- Response Time 100 ms
- Span Shift 0.02% FS/°C/atm
- Turndown Ratio 200:1
- Volumetric Flow Rate, Min 0.0025 SCCM
- Width 1.1 in
The FMA-LP2600A Series mass flow controllers use the principle of differential pressure within a laminar flow field to determine and control mass flow rate. A laminar flow element (LFE) inside the meter forces the gas into laminar (streamlined) flow. Inside this region, the Poiseuille equation dictates that the volumetric flow rate be linearly related to the pressure drop. A differential pressure sensor is used to measure the pressure drop along a fixed distance of the LFE. This, along with the viscosity of the gas, is used to accurately determine the volumetric flow rate. Separate absolute temperature and pressure sensors are incorporated and correct the volumetric flow rate to a set of standard conditions. This standardized flow rate is commonly called the mass flow rate and is reported in units such as standard cubic centimeters per minute (SCCM) or standard liters per minute (SLM).
The controller uses a true proportional valve coupled to the flow body to control flow using the integral PID loop controller. Standard units include a 0 to 5V output (4 to 20 mA optional) and RS232 communications. The gas-select feature and the set points can be adjusted from the front keypad or via RS232 communications. Volumetric flow, mass flow, absolute pressure, and temperature can all be viewed or recorded through the RS232 connection. It is also possible to multi-drop up to 26 units on the same serial connection to a distance of 46 m (150').
The controller uses a true proportional valve coupled to the flow body to control flow using the integral PID loop controller. Standard units include a 0 to 5V output (4 to 20 mA optional) and RS232 communications. The gas-select feature and the set points can be adjusted from the front keypad or via RS232 communications. Volumetric flow, mass flow, absolute pressure, and temperature can all be viewed or recorded through the RS232 connection. It is also possible to multi-drop up to 26 units on the same serial connection to a distance of 46 m (150').