Product Overview
- Accuracy Class > 0.25%
- Additional Features 1 8 Selectable Units
- Battery Size and Type Alkaline AAA
- Data Interface N/A
- Height 132.9 mm
- Length 24.7 mm
- Number of Batteries 2
- Number of Digits 5
- Operating Temperature, Max 50 °C
- Operating Temperature, Min 0 °C
- Pressure Resolution 0.1 hpa
- Supply Power Note 2 “AAA” batteries (included)
- Width 50.9 mm
The Basic Manometer HHC280 series features absolute and differential pressure measurement with a clear 5-digit display. Designed for precision, it includes an auto-off function to conserve battery life, making it suitable for various industrial applications requiring reliable pressure monitoring.
PDFs & Manuals
Show Ratings & Reviews
Rated 3 out of
5
by
Mazzerack from
Good Basic Instrument
I purchased this Manometer for laboratory testing on residential HVAC duct systems. I am measuring duct pressure "Inches/WC. The specifications stated the Instrument had an "averaging" feature which I badly needed. Turns out the "averaging" feature was only for wind speed, so that was very disappointing. But the instrument works very well and provides stable readings.
Date published: 2016-10-05
Rated 5 out of
5
by
IanW from
Well-priced, accurate, and arrived quickly!
It is almost impossible to find an absolute pressure gauge that can cover these ranges. It was relatively inexpensive, accurate, and I received it in just a couple of days after a very easy ordering process. Great product!
Date published: 2019-02-15
Would this unit be suitable for measuring pressure differentials across adjacent cleanrooms of a cleanroom suite? In particular, I want to use the absolute pressure unit because I don't want to drill holes in the walls separating the adjacent walls.
Yes. You can move the meter from room to room and monitor the change in the readings.
Date published: 2020-04-03
What is the resolution of HHC280 when measuring under 100 Pa? The datasheet is a bit confusing. Is it 10 Pa or 1Pa?
Thank you for your inquiry, the Resolution is : 0.1 hpa/0.1 mmH2O
Date published: 2022-02-08
What we can do in case of OOT ? Thanks in advance
If the HHC280 unit is out of tolerance it would need to be returned to us for evaluation. There are no field calibration procedures available.
Date published: 2021-06-14